CNA source text (English)
An issue was discovered in Smiths-Medical CADD-Solis Medication Safety Software, Version 1.0; 2.0; 3.0; and 3.1. CADD-Solis Medication Safety Software grants an authenticated user elevated privileges on the SQL database, which would allow an authenticated user to modify drug libraries, add and delete users, and change user permissions. According to Smiths-Medical, physical access to the pump is required to install drug library updates.
Scoped product and version facts
Fix information unknown; no fixed version is derived here. Missing information does not prove that no fix exists.
Exploitation unknown; this catalogue does not assess exploitation.
Smiths-Medical CADD-Solis Medication Safety Software through 3.1
- Vendor
- n/a
- Product
- Smiths-Medical CADD-Solis Medication Safety Software through 3.1
- Platforms
- Unknown
- Default status
Unknown
| Version (version) | Status (status) | Version scheme (versionType) | Exclusive end (lessThan) | Inclusive end (lessThanOrEqual) | Status changes (at → status) |
|---|---|---|---|---|---|
| Smiths-Medical CADD-Solis Medication Safety Software through 3.1 | affected | Unavailable | Unavailable | Unavailable | None listed |