CVE-2017-12726
CVE.org · Published
- Record status
- Published
- Identifier year (not publication year)
- 2017
- Source publication
- Source updated
- Imported by cveboard
- Source snapshot
- cve_2026-10-24_midnight_baseline ·
CNA source text (English)
A Use of Hard-coded Password issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. Telnet on the pump uses hardcoded credentials, which can be used if the pump is configured to allow external communications. Smiths Medical assesses that it is not possible to upload files via Telnet and the impact of this vulnerability is limited to the communications module.
Scoped product and version facts
Fix information unknown; no fixed version is derived here. Missing information does not prove that no fix exists.
Exploitation unknown; this catalogue does not assess exploitation.
Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump
- Vendor
- n/a
- Product
- Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump
- Platforms
- Unknown
- Default status
Unknown
| Version (version) | Status (status) | Version scheme (versionType) | Exclusive end (lessThan) | Inclusive end (lessThanOrEqual) | Status changes (at → status) |
|---|---|---|---|---|---|
| Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump | affected | Unavailable | Unavailable | Unavailable | None listed |