  "schemaVersion": 1,
  "attribution": [
    {
      "source": "CISA Known Exploited Vulnerabilities",
      "url": "https://www.cisa.gov/known-exploited-vulnerabilities-catalog",
      "licence": "US government work"
    },
    {
      "source": "GitHub Advisory Database",
      "url": "https://github.com/advisories",
      "licence": "CC BY 4.0"
    },
    {
      "source": "Have I Been Pwned",
      "url": "https://haveibeenpwned.com/",
      "licence": "CC BY 4.0"
    },
    {
      "source": "CVE Program (CVE records)",
      "url": "https://www.cve.org/Legal/TermsOfUse",
      "licence": "CVE Terms of Use",
      "notice": "Copyright © 1999-2026, The MITRE Corporation. CVE is a trademark and the CVE logo is a registered trademark of The MITRE Corporation.",
      "terms": "CVE Usage: MITRE hereby grants you a perpetual, worldwide, non-exclusive, no-charge, royalty-free, irrevocable copyright license to reproduce, prepare derivative works of, publicly display, publicly perform, sublicense, and distribute Common Vulnerabilities and Exposures (CVE™). Any copy you make for such purposes is authorized provided that you reproduce MITRE's copyright designation and this license in any such copy."
    },
    {
      "source": "FIRST EPSS",
      "url": "https://www.first.org/epss/",
      "licence": "free to use; FIRST requests attribution"
    }
  ],
  "record": {
    "id": "CVE-2018-10631",
    "sourceURL": "https://www.cve.org/CVERecord?id=CVE-2018-10631",
    "idYear": 2018,
    "status": "PUBLISHED",
    "publishedAt": "2018-07-13T19:00:00.000Z",
    "updatedAt": "2025-08-26T14:26:52.840Z",
    "importedAt": "2026-10-24T01:56:00.349Z",
    "snapshot": {
      "id": "cve_2026-10-24_midnight_baseline",
      "at": "2026-10-24T00:00:00.000Z",
      "source": "https://github.com/CVEProject/cvelistV5/releases/tag/cve_2026-10-24_0100Z"
    },
    "facts": {
      "title": "Medtronic N'Vision Clinician Programmer Protection Mechanism Failure",
      "summary": "The 8840 Clinician Programmer executes the application program from the 8870 Application Card. An attacker with physical access to an 8870 Application Card and sufficient technical capability can modify the contents of this card, including the binary executables. If modified to bypass protection mechanisms, this malicious code will be run when the card is inserted into an 8840 Clinician Programmer.",
      "affected": [
        {
          "vendor": "Medtronic",
          "product": "8840 N’Vision Clinician Programmer",
          "defaultStatus": "unaffected",
          "versions": [
            {
              "version": "All versions",
              "status": "affected"
            }
          ]
        },
        {
          "vendor": "Medtronic",
          "product": "8870 N’Vision removable Application Card",
          "defaultStatus": "unaffected",
          "versions": [
            {
              "version": "All versions",
              "status": "affected"
            }
          ]
        }
      ]
    }
  }
}