CVE-2018-19010
CVE.org · Published
- Record status
- Published
- Identifier year (not publication year)
- 2018
- Source publication
- Source updated
- Imported by cveboard
- Source snapshot
- cve_2026-10-24_midnight_baseline ·
CNA source text (English)
Drager Infinity Delta, Infinity Delta, all versions, Delta XL, all versions, Kappa, all version, and Infinity Explorer C700, all versions. A malformed network packet may cause the monitor to reboot. By repeatedly sending the malformed network packet, an attacker may be able to disrupt patient monitoring by causing the monitor to repeatedly reboot until it falls back to default configuration and loses network connectivity.
Scoped product and version facts
Fix information unknown; no fixed version is derived here. Missing information does not prove that no fix exists.
Exploitation unknown; this catalogue does not assess exploitation.
Dräger Infinity Delta
- Vendor
- ICS-CERT
- Product
- Dräger Infinity Delta
- Platforms
- Unknown
- Default status
Unknown
| Version (version) | Status (status) | Version scheme (versionType) | Exclusive end (lessThan) | Inclusive end (lessThanOrEqual) | Status changes (at → status) |
|---|---|---|---|---|---|
| Infinity Delta, all versions, Delta XL, all versions, Kappa, all version, and Infinity Explorer C700, all versions. | affected | Unavailable | Unavailable | Unavailable | None listed |