  "schemaVersion": 1,
  "attribution": [
    {
      "source": "CISA Known Exploited Vulnerabilities",
      "url": "https://www.cisa.gov/known-exploited-vulnerabilities-catalog",
      "licence": "US government work"
    },
    {
      "source": "GitHub Advisory Database",
      "url": "https://github.com/advisories",
      "licence": "CC BY 4.0"
    },
    {
      "source": "Have I Been Pwned",
      "url": "https://haveibeenpwned.com/",
      "licence": "CC BY 4.0"
    },
    {
      "source": "CVE Program (CVE records)",
      "url": "https://www.cve.org/Legal/TermsOfUse",
      "licence": "CVE Terms of Use",
      "notice": "Copyright © 1999-2026, The MITRE Corporation. CVE is a trademark and the CVE logo is a registered trademark of The MITRE Corporation.",
      "terms": "CVE Usage: MITRE hereby grants you a perpetual, worldwide, non-exclusive, no-charge, royalty-free, irrevocable copyright license to reproduce, prepare derivative works of, publicly display, publicly perform, sublicense, and distribute Common Vulnerabilities and Exposures (CVE™). Any copy you make for such purposes is authorized provided that you reproduce MITRE's copyright designation and this license in any such copy."
    },
    {
      "source": "FIRST EPSS",
      "url": "https://www.first.org/epss/",
      "licence": "free to use; FIRST requests attribution"
    }
  ],
  "record": {
    "id": "CVE-2018-8868",
    "sourceURL": "https://www.cve.org/CVERecord?id=CVE-2018-8868",
    "idYear": 2018,
    "status": "PUBLISHED",
    "publishedAt": "2018-07-02T18:00:00.000Z",
    "updatedAt": "2025-05-22T18:14:07.710Z",
    "importedAt": "2026-10-24T01:56:04.670Z",
    "snapshot": {
      "id": "cve_2026-10-24_midnight_baseline",
      "at": "2026-10-24T00:00:00.000Z",
      "source": "https://github.com/CVEProject/cvelistV5/releases/tag/cve_2026-10-24_0100Z"
    },
    "facts": {
      "title": "Medtronic MyCareLink Patient Monitor Exposed Dangerous Method or Function",
      "summary": "Medtronic 24950 MyCareLink Monitor and 24952 MyCareLink Monitor contains debug code meant to test the functionality of the monitor's communication interfaces, including the interface between the monitor and implantable cardiac device. An attacker with physical access to the device can exploit other vulnerabilities to access this debug functionality. This debug functionality provides the ability to read and write arbitrary memory values to implantable cardiac devices via inductive or short range wireless protocols. An attacker with close physical proximity to a target implantable cardiac device can use this debug functionality.",
      "affected": [
        {
          "vendor": "Medtronic",
          "product": "24950 MyCareLink Monitor",
          "defaultStatus": "unaffected",
          "versions": [
            {
              "version": "All versions",
              "status": "affected"
            }
          ]
        },
        {
          "vendor": "Medtronic",
          "product": "24952 MyCareLink Monitor",
          "defaultStatus": "unaffected",
          "versions": [
            {
              "version": "All versions",
              "status": "affected"
            }
          ]
        }
      ]
    }
  }
}