CVE-2019-25723
CVE.org · Published
- Record status
- Published
- Identifier year (not publication year)
- 2019
- Source publication
- Source updated
- Imported by cveboard
- Source snapshot
- cve_2026-10-24_midnight_baseline ·
CNA source text (English)
Dräger Perseus A500 2.00-2.02 DoS via Medibus Interface
Dräger Perseus A500 software versions 2.00 through 2.02 contains an improper input handling vulnerability that allows external attackers to cause a denial of service by sending specifically crafted non-Medibus-compliant data through the Medibus interface. Attackers can overload the internal processor with malformed data to trigger a warm restart, causing ventilation pressure to drop to ambient level and interrupting ventilation for several seconds before therapy resumes.
Scoped product and version facts
Fix information unknown; no fixed version is derived here. Missing information does not prove that no fix exists.
Exploitation unknown; this catalogue does not assess exploitation.
Perseus A500
- Vendor
- Dräger
- Product
- Perseus A500
- Platforms
- Unknown
- Default status
unknown
| Version (version) | Status (status) | Version scheme (versionType) | Exclusive end (lessThan) | Inclusive end (lessThanOrEqual) | Status changes (at → status) |
|---|---|---|---|---|---|
| 2.00 | affected | custom | Unavailable | Unavailable | None listed |
| 2.01 | affected | custom | Unavailable | Unavailable | None listed |
| 2.02 | affected | custom | Unavailable | Unavailable | None listed |
| 2.03 | unaffected | custom | Unavailable | Unavailable | None listed |