CNA source text (English)

Dräger Infinity M300 VG2.x Network-Based Denial of Service

Dräger Infinity M300 patient worn monitors with software version VG2.x and earlier contain a network-based denial of service vulnerability that allows attackers with access to the hospital or Infinity Network to repeatedly trigger device reboots until the device enters a fail state requiring manual restart. Attackers can exploit this vulnerability to cause loss of wireless network connectivity, temporary loss of patient monitoring, and interruption of alarm functionality until the device is manually recovered.

Scoped product and version facts

Fix information unknown; no fixed version is derived here. Missing information does not prove that no fix exists.

Exploitation unknown; this catalogue does not assess exploitation.

Infinity M300

Vendor
Dräger
Product
Infinity M300
Platforms
Unknown
Default status
unknown
Infinity M300 · Scoped product and version facts
Version (version)Status (status)Version scheme (versionType)Exclusive end (lessThan)Inclusive end (lessThanOrEqual)Status changes (at → status)
VG2.xaffectedcustomUnavailableVG2.3.1None listed
VG3.0unaffectedcustomUnavailableUnavailableNone listed