CVE-2020-12040
CVE.org · Published
- Record status
- Published
- Identifier year (not publication year)
- 2020
- Source publication
- Source updated
- Imported by cveboard
- Source snapshot
- cve_2026-10-24_midnight_baseline ·
CNA source text (English)
Sigma Spectrum Infusion System v's6.x (model 35700BAX) and Baxter Spectrum Infusion System Version(s) 8.x (model 35700BAX2) at the application layer uses an unauthenticated clear-text communication channel to send and receive system status and operational data. This could allow an attacker that has circumvented network security measures to view sensitive non-private data or to perform a man-in-the-middle attack.
Scoped product and version facts
Fix information unknown; no fixed version is derived here. Missing information does not prove that no fix exists.
Exploitation unknown; this catalogue does not assess exploitation.
Baxter Sigma Spectrum Infusion Pumps
- Vendor
- n/a
- Product
- Baxter Sigma Spectrum Infusion Pumps
- Platforms
- Unknown
- Default status
Unknown
| Version (version) | Status (status) | Version scheme (versionType) | Exclusive end (lessThan) | Inclusive end (lessThanOrEqual) | Status changes (at → status) |
|---|---|---|---|---|---|
| Sigma Spectrum v6.x model 35700BAX, Baxter Spectrum v8.x model 35700BAX2,Sigma Spectrum v6.x with Wireless Battery Module v9,11,13,14,15,16,v20D29,v20D30,v20D31,v22D24, Baxter Spectrum v8.x with Wireless Battery Module v17,v20D29,v20D30,v20D31,v22D24,Baxter Spectrum Wireless Battery Module v17,v20D29,v20D30,v20D31,v22D24,Baxter Spectrum LVP v8.x w/Wireless Battery Module v17,v20D29,v20D30,v20D31,v22D24 | affected | Unavailable | Unavailable | Unavailable | None listed |