  "schemaVersion": 1,
  "attribution": [
    {
      "source": "CISA Known Exploited Vulnerabilities",
      "url": "https://www.cisa.gov/known-exploited-vulnerabilities-catalog",
      "licence": "US government work"
    },
    {
      "source": "GitHub Advisory Database",
      "url": "https://github.com/advisories",
      "licence": "CC BY 4.0"
    },
    {
      "source": "Have I Been Pwned",
      "url": "https://haveibeenpwned.com/",
      "licence": "CC BY 4.0"
    },
    {
      "source": "CVE Program (CVE records)",
      "url": "https://www.cve.org/Legal/TermsOfUse",
      "licence": "CVE Terms of Use",
      "notice": "Copyright © 1999-2026, The MITRE Corporation. CVE is a trademark and the CVE logo is a registered trademark of The MITRE Corporation.",
      "terms": "CVE Usage: MITRE hereby grants you a perpetual, worldwide, non-exclusive, no-charge, royalty-free, irrevocable copyright license to reproduce, prepare derivative works of, publicly display, publicly perform, sublicense, and distribute Common Vulnerabilities and Exposures (CVE™). Any copy you make for such purposes is authorized provided that you reproduce MITRE's copyright designation and this license in any such copy."
    },
    {
      "source": "FIRST EPSS",
      "url": "https://www.first.org/epss/",
      "licence": "free to use; FIRST requests attribution"
    }
  ],
  "record": {
    "id": "CVE-2020-16220",
    "sourceURL": "https://www.cve.org/CVERecord?id=CVE-2020-16220",
    "idYear": 2020,
    "status": "PUBLISHED",
    "publishedAt": "2020-09-11T13:11:22.000Z",
    "updatedAt": "2024-08-04T13:37:53.493Z",
    "importedAt": "2026-10-24T01:56:17.081Z",
    "snapshot": {
      "id": "cve_2026-10-24_midnight_baseline",
      "at": "2026-10-24T00:00:00.000Z",
      "source": "https://github.com/CVEProject/cvelistV5/releases/tag/cve_2026-10-24_0100Z"
    },
    "facts": {
      "title": "Philips Patient Monitoring Devices Improper Validation of Syntactic Correctness of Input",
      "summary": "In Patient Information Center iX (PICiX) Versions C.02, C.03, \nPerformanceBridge Focal Point Version A.01, the product receives input \nthat is expected to be well-formed (i.e., to comply with a certain \nsyntax) but it does not validate or incorrectly validates that the input\n complies with the syntax, causing the certificate enrollment service to\n crash. It does not impact monitoring but prevents new devices from \nenrolling.",
      "affected": [
        {
          "vendor": "Philips",
          "product": "Patient Information Center iX (PICiX)",
          "defaultStatus": "unaffected",
          "versions": [
            {
              "version": "C.02",
              "status": "affected"
            },
            {
              "version": "C.03",
              "status": "affected"
            }
          ]
        },
        {
          "vendor": "Philips",
          "product": "PerformanceBridge Focal Point",
          "defaultStatus": "unaffected",
          "versions": [
            {
              "version": "A.01",
              "status": "affected"
            }
          ]
        }
      ]
    }
  }
}