cveboard RSS feed

CVE-2020-27252

CVE.org · Published

Historical CVEsJSONAI instructions

Record status
Published
Identifier year (not publication year)
2020
Source publication
Source updated
Imported by cveboard
Source snapshot
cve_2026-10-24_midnight_baseline ·

Plain text

CNA source text (English)

Medtronic MyCareLink Smart Time-of-check Time-of-use Race Condition

Medtronic MyCareLink Smart 25000 is vulnerable to a race condition in the MCL Smart Patient Reader software update system, which allows unsigned firmware to be uploaded and executed on the Patient Reader. If exploited, an attacker could remotely execute code on the MCL Smart Patient Reader device, leading to control of the device.

Scoped product and version facts

Fix information unknown; no fixed version is derived here. Missing information does not prove that no fix exists.

Exploitation unknown; this catalogue does not assess exploitation.

Smart Model 25000 Patient Reader

Vendor
Medtronic
Product
Smart Model 25000 Patient Reader
Platforms
Unknown
Default status
unaffected
Smart Model 25000 Patient Reader · Scoped product and version facts
Version (version)Status (status)Version scheme (versionType)Exclusive end (lessThan)Inclusive end (lessThanOrEqual)Status changes (at → status)
All versionsaffectedUnavailableUnavailableUnavailableNone listed