CVE-2022-38100
CVE.org · Published
- Record status
- Published
- Identifier year (not publication year)
- 2022
- Source publication
- Source updated
- Imported by cveboard
- Source snapshot
- cve_2026-10-24_midnight_baseline ·
CNA source text (English)
Contec Health CMS8000
The CMS800 device fails while attempting to parse malformed network data sent by a threat actor. A threat actor with network access can remotely issue a specially formatted UDP request that will cause the entire device to crash and require a physical reboot. A UDP broadcast request could be sent that causes a mass denial-of-service attack on all CME8000 devices connected to the same network.
Scoped product and version facts
Fix information unknown; no fixed version is derived here. Missing information does not prove that no fix exists.
Exploitation unknown; this catalogue does not assess exploitation.
CMS8000 CONTEC ICU CCU Vital Signs Patient Monitor
- Vendor
- Contec Health
- Product
- CMS8000 CONTEC ICU CCU Vital Signs Patient Monitor
- Platforms
- Unknown
- Default status
Unknown
| Version (version) | Status (status) | Version scheme (versionType) | Exclusive end (lessThan) | Inclusive end (lessThanOrEqual) | Status changes (at → status) |
|---|---|---|---|---|---|
| All | affected | Unavailable | Unavailable | Unavailable | None listed |