  "schemaVersion": 1,
  "attribution": [
    {
      "source": "CISA Known Exploited Vulnerabilities",
      "url": "https://www.cisa.gov/known-exploited-vulnerabilities-catalog",
      "licence": "US government work"
    },
    {
      "source": "GitHub Advisory Database",
      "url": "https://github.com/advisories",
      "licence": "CC BY 4.0"
    },
    {
      "source": "Have I Been Pwned",
      "url": "https://haveibeenpwned.com/",
      "licence": "CC BY 4.0"
    },
    {
      "source": "CVE Program (CVE records)",
      "url": "https://www.cve.org/Legal/TermsOfUse",
      "licence": "CVE Terms of Use",
      "notice": "Copyright © 1999-2026, The MITRE Corporation. CVE is a trademark and the CVE logo is a registered trademark of The MITRE Corporation.",
      "terms": "CVE Usage: MITRE hereby grants you a perpetual, worldwide, non-exclusive, no-charge, royalty-free, irrevocable copyright license to reproduce, prepare derivative works of, publicly display, publicly perform, sublicense, and distribute Common Vulnerabilities and Exposures (CVE™). Any copy you make for such purposes is authorized provided that you reproduce MITRE's copyright designation and this license in any such copy."
    },
    {
      "source": "FIRST EPSS",
      "url": "https://www.first.org/epss/",
      "licence": "free to use; FIRST requests attribution"
    }
  ],
  "record": {
    "id": "CVE-2022-4992",
    "sourceURL": "https://www.cve.org/CVERecord?id=CVE-2022-4992",
    "idYear": 2022,
    "status": "PUBLISHED",
    "publishedAt": "2026-06-02T21:06:40.281Z",
    "updatedAt": "2026-06-05T11:26:40.821Z",
    "importedAt": "2026-10-24T01:56:49.724Z",
    "snapshot": {
      "id": "cve_2026-10-24_midnight_baseline",
      "at": "2026-10-24T00:00:00.000Z",
      "source": "https://github.com/CVEProject/cvelistV5/releases/tag/cve_2026-10-24_0100Z"
    },
    "facts": {
      "title": "Dräger Infinity M540 VG4.1.1 Spoofed Network Message Handling DoS/Tampering",
      "summary": "Dräger Infinity Acute Care System and Standalone Infinity M540 patient monitors versions VG4.1.1, VG4.0.3, and lower (with VG4.2 partially affected) contain a network message handling vulnerability that allows remote attackers to inject spoofed or tampered data and cause denial-of-service conditions. Attackers can compromise network communications to modify device settings such as alarm states or alarm limits, or overwhelm the system with excessive network traffic causing the Cockpit or M540 to reboot and lose network functionality.",
      "affected": [
        {
          "vendor": "Dräger",
          "product": "Infinity Acute Care System",
          "defaultStatus": "unknown",
          "versions": [
            {
              "version": "0",
              "status": "affected",
              "versionType": "custom",
              "lessThan": "VG4.2"
            },
            {
              "version": "0",
              "status": "affected",
              "versionType": "custom",
              "lessThan": "VG4.1.1"
            },
            {
              "version": "0",
              "status": "affected",
              "versionType": "custom",
              "lessThan": "VG4.0.3"
            }
          ]
        },
        {
          "vendor": "Dräger",
          "product": "Standalone Infinity M540 patient monitor",
          "defaultStatus": "unknown",
          "versions": [
            {
              "version": "0",
              "status": "affected",
              "versionType": "custom",
              "lessThan": "VG4.2"
            },
            {
              "version": "0",
              "status": "affected",
              "versionType": "custom",
              "lessThan": "VG4.1.1"
            }
          ]
        }
      ]
    }
  }
}