CVE-2023-0888
CVE.org · Published
- Record status
- Published
- Identifier year (not publication year)
- 2023
- Source publication
- Source updated
- Imported by cveboard
- Source snapshot
- cve_2026-10-24_midnight_baseline ·
CNA source text (English)
Authenticated eval injection in B. Braun Space Battery pack SP with Wi-Fi
An improper neutralization of directives in dynamically evaluated code vulnerability in the WiFi Battery embedded web server in versions L90/U70 and L92/U92 can be used to gain administrative access to the WiFi communication module. An authenticated user, having access to both the medical device WiFi network (such as a biomedical engineering staff member) and the specific B.Braun Battery Pack SP with WiFi web server credentials, could get administrative (root) access on the infusion pump communication module. This could be used as a vector to start further attacks
Scoped product and version facts
Fix information unknown; no fixed version is derived here. Missing information does not prove that no fix exists.
Exploitation unknown; this catalogue does not assess exploitation.
Space Battery Pack SP with Wi-Fi
- Vendor
- B. Braun
- Product
- Space Battery Pack SP with Wi-Fi
- Platforms
- Unknown
- Default status
unaffected
| Version (version) | Status (status) | Version scheme (versionType) | Exclusive end (lessThan) | Inclusive end (lessThanOrEqual) | Status changes (at → status) |
|---|---|---|---|---|---|
| 0 | affected | custom | Unavailable | 053L000092 | None listed |
| 0 | affected | custom | Unavailable | 054U000092 | None listed |