CNA source text (English)
Life2000 service tools for test and calibration do not support user authentication
The software tools used by service personnel to test & calibrate the ventilator do not support user authentication. An attacker with access to the Service PC where the tools are installed could obtain diagnostic information through the test tool or manipulate the ventilator's settings and embedded software via the calibration tool, without having to authenticate to either tool. This could result in unauthorized disclosure of information and/or have unintended impacts on device settings and performance.
Scoped product and version facts
Fix information unknown; no fixed version is derived here. Missing information does not prove that no fix exists.
Exploitation unknown; this catalogue does not assess exploitation.
Life2000 Ventilation System
- Vendor
- Baxter
- Product
- Life2000 Ventilation System
- Platforms
- Unknown
- Default status
unaffected
| Version (version) | Status (status) | Version scheme (versionType) | Exclusive end (lessThan) | Inclusive end (lessThanOrEqual) | Status changes (at → status) |
|---|---|---|---|---|---|
| 06.08.00.00 and prior | affected | Unavailable | Unavailable | Unavailable | None listed |