CVE-2024-48971
CVE.org · Published
- Record status
- Published
- Identifier year (not publication year)
- 2024
- Source publication
- Source updated
- Imported by cveboard
- Source snapshot
- cve_2026-10-24_midnight_baseline ·
CNA source text (English)
Clinician Password and Serial Number Clinician Password are hard-coded in Life2000 Ventilator
The Clinician Password and Serial Number Clinician Password are hard-coded into the ventilator in plaintext form. This could allow an attacker to obtain the password off the ventilator and use it to gain unauthorized access to the device, with clinician privileges.
Scoped product and version facts
Fix information unknown; no fixed version is derived here. Missing information does not prove that no fix exists.
Exploitation unknown; this catalogue does not assess exploitation.
Life2000 Ventilation System
- Vendor
- Baxter
- Product
- Life2000 Ventilation System
- Platforms
- Unknown
- Default status
unaffected
| Version (version) | Status (status) | Version scheme (versionType) | Exclusive end (lessThan) | Inclusive end (lessThanOrEqual) | Status changes (at → status) |
|---|---|---|---|---|---|
| 06.08.00.00 and prior | affected | Unavailable | Unavailable | Unavailable | None listed |