  "schemaVersion": 1,
  "attribution": [
    {
      "source": "CISA Known Exploited Vulnerabilities",
      "url": "https://www.cisa.gov/known-exploited-vulnerabilities-catalog",
      "licence": "US government work"
    },
    {
      "source": "GitHub Advisory Database",
      "url": "https://github.com/advisories",
      "licence": "CC BY 4.0"
    },
    {
      "source": "Have I Been Pwned",
      "url": "https://haveibeenpwned.com/",
      "licence": "CC BY 4.0"
    },
    {
      "source": "CVE Program (CVE records)",
      "url": "https://www.cve.org/Legal/TermsOfUse",
      "licence": "CVE Terms of Use",
      "notice": "Copyright © 1999-2026, The MITRE Corporation. CVE is a trademark and the CVE logo is a registered trademark of The MITRE Corporation.",
      "terms": "CVE Usage: MITRE hereby grants you a perpetual, worldwide, non-exclusive, no-charge, royalty-free, irrevocable copyright license to reproduce, prepare derivative works of, publicly display, publicly perform, sublicense, and distribute Common Vulnerabilities and Exposures (CVE™). Any copy you make for such purposes is authorized provided that you reproduce MITRE's copyright designation and this license in any such copy."
    },
    {
      "source": "FIRST EPSS",
      "url": "https://www.first.org/epss/",
      "licence": "free to use; FIRST requests attribution"
    }
  ],
  "record": {
    "id": "CVE-2024-48973",
    "sourceURL": "https://www.cve.org/CVERecord?id=CVE-2024-48973",
    "idYear": 2024,
    "status": "PUBLISHED",
    "publishedAt": "2024-11-14T21:24:14.099Z",
    "updatedAt": "2024-11-18T15:33:44.924Z",
    "importedAt": "2026-10-24T01:57:16.009Z",
    "snapshot": {
      "id": "cve_2026-10-24_midnight_baseline",
      "at": "2026-10-24T00:00:00.000Z",
      "source": "https://github.com/CVEProject/cvelistV5/releases/tag/cve_2026-10-24_0100Z"
    },
    "facts": {
      "title": "Debug port on Life2000 Ventilator serial interface is enabled by default",
      "summary": "The debug port on the ventilator's serial interface is enabled by default. This could allow an attacker to send and receive messages over the debug port (which are unencrypted; see 3.2.1) that result in unauthorized disclosure of information and/or have unintended impacts on device settings and performance.",
      "affected": [
        {
          "vendor": "Baxter",
          "product": "Life2000 Ventilation System",
          "defaultStatus": "unaffected",
          "versions": [
            {
              "version": "06.08.00.00 and prior",
              "status": "affected"
            }
          ]
        }
      ]
    }
  }
}