CVE-2024-9832
CVE.org · Published
- Record status
- Published
- Identifier year (not publication year)
- 2024
- Source publication
- Source updated
- Imported by cveboard
- Source snapshot
- cve_2026-10-24_midnight_baseline ·
CNA source text (English)
No limit on failed login attempts with Clinician Password or Serial Number Clinician Password on Life2000 Ventilator
There is no limit on the number of failed login attempts permitted with the Clinician Password or the Serial Number Clinician Password. An attacker could execute a brute-force attack to gain unauthorized access to the ventilator, and then make changes to device settings that could disrupt the function of the device and/or result in unauthorized information disclosure.
Scoped product and version facts
Fix information unknown; no fixed version is derived here. Missing information does not prove that no fix exists.
Exploitation unknown; this catalogue does not assess exploitation.
Life2000 Ventilation System
- Vendor
- Baxter
- Product
- Life2000 Ventilation System
- Platforms
- Unknown
- Default status
unaffected
| Version (version) | Status (status) | Version scheme (versionType) | Exclusive end (lessThan) | Inclusive end (lessThanOrEqual) | Status changes (at → status) |
|---|---|---|---|---|---|
| 06.08.00.00 and prior | affected | Unavailable | Unavailable | Unavailable | None listed |