cveboard RSS feed

CVE-2025-0626

CVE.org · Published

Historical CVEsJSONAI instructions

Record status
Published
Identifier year (not publication year)
2025
Source publication
Source updated
Imported by cveboard
Source snapshot
cve_2026-10-24_midnight_baseline ·

Plain text

CNA source text (English)

Hidden Functionality vulnerability in Contec Health CMS8000 Patient Monitor

The "monitor" binary in the firmware of the affected product attempts to mount to a hard-coded, routable IP address, bypassing existing device network settings to do so. The function also enables the network interface of the device if it is disabled. The function is triggered by attempting to update the device from the user menu. This could serve as a backdoor to the device, and could lead to a malicious actor being able to upload and overwrite files on the device.

Scoped product and version facts

Fix information unknown; no fixed version is derived here. Missing information does not prove that no fix exists.

Exploitation unknown; this catalogue does not assess exploitation.

CMS8000 Patient Monitor

Vendor
Contec Health
Product
CMS8000 Patient Monitor
Platforms
Unknown
Default status
unaffected
CMS8000 Patient Monitor · Scoped product and version facts
Version (version)Status (status)Version scheme (versionType)Exclusive end (lessThan)Inclusive end (lessThanOrEqual)Status changes (at → status)
All versionsaffectedUnavailableUnavailableUnavailableNone listed