CVE-2025-1204
CVE.org · Published
- Record status
- Published
- Identifier year (not publication year)
- 2025
- Source publication
- Source updated
- Imported by cveboard
- Source snapshot
- cve_2026-10-24_midnight_baseline ·
CNA source text (English)
The "update" binary in the firmware of the affected product sends attempts to mount to a hard-coded, routable IP address, bypassing existing device network settings to do so. The function triggers if the 'C' button is pressed at a specific time during the boot process. If an attacker is able to control or impersonate this IP address, they could upload and overwrite files on the device.
Scoped product and version facts
Fix information unknown; no fixed version is derived here. Missing information does not prove that no fix exists.
Exploitation unknown; this catalogue does not assess exploitation.
CMS8000 Patient Monitor
- Vendor
- Contec Health
- Product
- CMS8000 Patient Monitor
- Platforms
- Unknown
- Default status
unaffected
| Version (version) | Status (status) | Version scheme (versionType) | Exclusive end (lessThan) | Inclusive end (lessThanOrEqual) | Status changes (at → status) |
|---|---|---|---|---|---|
| 0 | affected | Unavailable | Unavailable | Unavailable | None listed |