CVE-2025-14095
CVE.org · Published
- Record status
- Published
- Identifier year (not publication year)
- 2025
- Source publication
- Source updated
- Imported by cveboard
- Source snapshot
- cve_2026-10-24_midnight_baseline ·
CNA source text (English)
Privilege boundary violation in Radiometer Products
A "Privilege boundary violation" vulnerability is identified affecting multiple Radiometer Products. Exploitation of this vulnerability gives a user with physical access to the analyzer, the possibility to gain unauthorized access to functionalities outside the restricted environment. The vulnerability is due to weakness in the design of access control implementation in application software. Other related CVE's are CVE-2025-14096 & CVE-2025-14097. Affected customers have been informed about this vulnerability. This CVE is being published to provide transparency. Required configuration for Exposure: Physical access to the analyzer is needed. Temporary work Around: Only authorized people can physically access the analyzer. Permanent solution: Local Radiometer representatives will contact all affected customers to discuss a permanent solution. Exploit Status: Researchers have provided working proof-of-concept. Radiometer is not aware of any publicly available exploit at the time of publication. Note: CVSS score 6.8 when underlying OS is Windows 7 or Windows XP Operating systems and CVSS score 5.7 when underlying OS is Windows 8 or Windows 10 operating systems.
Scoped product and version facts
Fix information unknown; no fixed version is derived here. Missing information does not prove that no fix exists.
Exploitation unknown; this catalogue does not assess exploitation.
ABL90 FLEX and ABL90 FLEX PLUS Analyzers
- Vendor
- Radiometer Medical Aps
- Product
- ABL90 FLEX and ABL90 FLEX PLUS Analyzers
- Platforms
- Unknown
- Default status
unaffected
| Version (version) | Status (status) | Version scheme (versionType) | Exclusive end (lessThan) | Inclusive end (lessThanOrEqual) | Status changes (at → status) |
|---|---|---|---|---|---|
| All application software versions with Windows 7, Windows XP as underlying OS | affected | ABL90 DMS(Data Management System) Application | Unavailable | Unavailable | None listed |
| Application software versions < 3.5MR11 with Windows 10 as underlying OS | affected | Unavailable | Unavailable | Unavailable | None listed |
AQT90 FLEX Analyzers
- Vendor
- Radiometer Medical Aps
- Product
- AQT90 FLEX Analyzers
- Platforms
- Unknown
- Default status
unaffected
| Version (version) | Status (status) | Version scheme (versionType) | Exclusive end (lessThan) | Inclusive end (lessThanOrEqual) | Status changes (at → status) |
|---|---|---|---|---|---|
| All Application software versions <= 8.13 MR2 | affected | AQT90 DMS(Data Management System) Application | Unavailable | Unavailable | None listed |
ABL800 BASIC and ABL800 FLEX Analyzers
- Vendor
- Radiometer Medical Aps
- Product
- ABL800 BASIC and ABL800 FLEX Analyzers
- Platforms
- Unknown
- Default status
unaffected
| Version (version) | Status (status) | Version scheme (versionType) | Exclusive end (lessThan) | Inclusive end (lessThanOrEqual) | Status changes (at → status) |
|---|---|---|---|---|---|
| Application software versions < 6.20 MR2 with Windows 7, Windows XP as underlying OS | affected | ABL800 DMS(Data Management System) | Unavailable | Unavailable | None listed |
| Application software versions < 6.20 MR2 with Windows 10 as underlying OS | affected | ABL800 DMS(Data Management System) | Unavailable | Unavailable | None listed |
ABL9 Analyzers
- Vendor
- Radiometer Medical Aps
- Product
- ABL9 Analyzers
- Platforms
- Unknown
- Default status
unaffected
| Version (version) | Status (status) | Version scheme (versionType) | Exclusive end (lessThan) | Inclusive end (lessThanOrEqual) | Status changes (at → status) |
|---|---|---|---|---|---|
| Application software versions < 1.5.0 | affected | CABO application | Unavailable | Unavailable | None listed |