  "schemaVersion": 1,
  "attribution": [
    {
      "source": "CISA Known Exploited Vulnerabilities",
      "url": "https://www.cisa.gov/known-exploited-vulnerabilities-catalog",
      "licence": "US government work"
    },
    {
      "source": "GitHub Advisory Database",
      "url": "https://github.com/advisories",
      "licence": "CC BY 4.0"
    },
    {
      "source": "Have I Been Pwned",
      "url": "https://haveibeenpwned.com/",
      "licence": "CC BY 4.0"
    },
    {
      "source": "CVE Program (CVE records)",
      "url": "https://www.cve.org/Legal/TermsOfUse",
      "licence": "CVE Terms of Use",
      "notice": "Copyright © 1999-2026, The MITRE Corporation. CVE is a trademark and the CVE logo is a registered trademark of The MITRE Corporation.",
      "terms": "CVE Usage: MITRE hereby grants you a perpetual, worldwide, non-exclusive, no-charge, royalty-free, irrevocable copyright license to reproduce, prepare derivative works of, publicly display, publicly perform, sublicense, and distribute Common Vulnerabilities and Exposures (CVE™). Any copy you make for such purposes is authorized provided that you reproduce MITRE's copyright designation and this license in any such copy."
    },
    {
      "source": "FIRST EPSS",
      "url": "https://www.first.org/epss/",
      "licence": "free to use; FIRST requests attribution"
    }
  ],
  "record": {
    "id": "CVE-2025-14097",
    "sourceURL": "https://www.cve.org/CVERecord?id=CVE-2025-14097",
    "idYear": 2025,
    "status": "PUBLISHED",
    "publishedAt": "2025-12-17T12:36:03.779Z",
    "updatedAt": "2025-12-17T14:41:13.252Z",
    "importedAt": "2026-10-24T01:57:24.991Z",
    "snapshot": {
      "id": "cve_2026-10-24_midnight_baseline",
      "at": "2026-10-24T00:00:00.000Z",
      "source": "https://github.com/CVEProject/cvelistV5/releases/tag/cve_2026-10-24_0100Z"
    },
    "facts": {
      "title": "Remote Code Execution Vulnerability in Radiometer Products",
      "summary": "A vulnerability in the application software of multiple Radiometer products may allow remote code execution and unauthorized device management when specific internal conditions are met. Exploitation requires that a remote connection is established with additional information obtained through other means.  The issue is caused by a weakness in the analyzer’s application software.                                                                                                                                                                                                Other related CVE's are CVE-2025-14095 & CVE-2025-14096.                                                                                                      Affected customers have been informed about this vulnerability. This CVE is being published to provide transparency.\n\n\n\nRequired Configuration for Exposure: Affected application software version is in use and remote support feature is enabled in the analyzer.                                                                                                                                                                        Temporary work Around: If the network is not considered secure, please remove the analyzer from the network.                         Permanent solution:\nCustomers should ensure the following:\n•  The network is secure, and access follows best practices.\nLocal Radiometer representatives will contact all affected customers to discuss a permanent solution.                                                     \nExploit Status:\n\n\nResearchers have provided working proof-of-concept (PoC). Radiometer is not aware of any publicly available exploits at the time of this publication.",
      "affected": [
        {
          "vendor": "Radiometer Medical Aps",
          "product": "ABL90 FLEX and ABL90 FLEX PLUS Analyzers",
          "defaultStatus": "unaffected",
          "versions": [
            {
              "version": "Application software versions < 3.5MR11 with Windows 7, Windows XP operating systems",
              "status": "affected",
              "versionType": "ABL90 DMS (Data Management system)"
            },
            {
              "version": "Application software versions >= 3.5MR11 with Windows 10 operating system",
              "status": "unaffected",
              "versionType": "ABL90 DMS (Data Management system)"
            }
          ]
        },
        {
          "vendor": "Radiometer Medical Aps",
          "product": "ABL800 BASIC and ABL800 FLEX Analyzers",
          "defaultStatus": "unaffected",
          "versions": [
            {
              "version": "Application software versions < 6.20MR2 with Windows 7, Windows XP operating systems",
              "status": "affected",
              "versionType": "ABL800 DMS(Data Management System)"
            },
            {
              "version": "Application software versions >= 6.20MR2 with Windows 10 operating system",
              "status": "unaffected",
              "versionType": "ABL800 DMS(Data Management System)"
            }
          ]
        },
        {
          "vendor": "Radiometer Medical Aps",
          "product": "AQT90 FLEX Analyzers",
          "defaultStatus": "unaffected",
          "versions": [
            {
              "version": "Application software versions <= 8.13 MR2 with Windows 7, Windows XP operating systems",
              "status": "affected",
              "versionType": "AQT90 DMS (Data Management System)"
            },
            {
              "version": "Application software versions >= 8.13 MR2 with Windows 10 operating system",
              "status": "unaffected",
              "versionType": "AQT90 DMS (Data Management System)"
            }
          ]
        }
      ]
    }
  }
}