cveboard RSS feed

CVE-2026-14297

CVE.org · Published

Historical CVEsJSONAI instructions

Record status
Published
Identifier year (not publication year)
2026
Source publication
Source updated
Imported by cveboard
Source snapshot
cve_2026-10-24_midnight_baseline ·

Plain text

CNA source text (English)

The Continuous Glucose Monitoring Service's Record Access Control Point (RACP) write handler `memcpy`s the entire attacker-supplied ATT write value into a fixed 20-byte BSS buffer.

A buffer overflow in the Bluetooth Continuous Glucose Monitoring Service (CGMS) Record Access Control Point (RACP) write handler allows an authenticated BLE peer to overflow a 20-byte static buffer into adjacent BSS memory. The exploitable impact cannot be predetermined - it is entirely dependent on the linker-assigned BSS layout of the specific firmware build, which may vary.

Scoped product and version facts

Fix information unknown; no fixed version is derived here. Missing information does not prove that no fix exists.

Exploitation unknown; this catalogue does not assess exploitation.

nRF Connect SDK

Vendor
Nordic Semiconductor ASA
Product
nRF Connect SDK
Platforms
Unknown
Default status
unaffected
nRF Connect SDK · Scoped product and version facts
Version (version)Status (status)Version scheme (versionType)Exclusive end (lessThan)Inclusive end (lessThanOrEqual)Status changes (at → status)
2.2.0affectedsemverUnavailable3.3.0None listed