CVE-2014-5431
CVE.org · Publicado
- Status do registro
- Publicado
- Ano do identificador (não é o ano de publicação)
- 2014
- Publicação na fonte
- Atualização na fonte
- Importado pelo cveboard
- Snapshot da fonte
- cve_2026-10-24_midnight_baseline ·
Texto da CNA em inglês
Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16 contains a hard-coded password, which provides access to basic biomedical information, limited device settings, and network configuration of the WBM, if connected. The hard-coded password may allow an attacker with physical access to the device to access management functions to make unauthorized configuration changes to biomedical settings such as turn on and off wireless connections and the phase-complete audible alarm that indicates the end of an infusion phase. Baxter has released a new version of the SIGMA Spectrum Infusion System, version 8, which incorporates hardware and software changes.
Fatos de produtos e versões por escopo
Dados de correção desconhecidos; nenhuma versão corrigida é deduzida aqui. Dados ausentes não comprovam ausência de correção.
Exploração desconhecida; este catálogo não avalia exploração.
SIGMA Spectrum Infusion System
- Fornecedor
- Baxter
- Produto
- SIGMA Spectrum Infusion System
- Plataformas
- Desconhecido
- Status padrão
Desconhecido
| Versão (version) | Status (status) | Esquema de versões (versionType) | Fim exclusivo (lessThan) | Fim inclusivo (lessThanOrEqual) | Mudanças de status (at → status) |
|---|---|---|---|---|---|
| 6.05 (model 35700BAX) with wireless battery module (WBM) version 16 | affected | Indisponível | Indisponível | Indisponível | Nenhuma listada |