cveboard RSS feed

CVE-2017-12725

CVE.org · Published

Historical CVEsJSONAI instructions

Record status
Published
Identifier year (not publication year)
2017
Source publication
Source updated
Imported by cveboard
Source snapshot
cve_2026-10-24_midnight_baseline ·

Plain text

CNA source text (English)

A Use of Hard-coded Credentials issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. The pump with default network configuration uses hard-coded credentials to automatically establish a wireless network connection. The pump will establish a wireless network connection even if the pump is Ethernet connected and active; however, if the wireless association is established and the Ethernet cable is attached, the pump does not attach the network stack to the wireless network. In this scenario, all network traffic is instead directed over the wired Ethernet connection.

Scoped product and version facts

Fix information unknown; no fixed version is derived here. Missing information does not prove that no fix exists.

Exploitation unknown; this catalogue does not assess exploitation.

Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump

Vendor
n/a
Product
Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump
Platforms
Unknown
Default status
Unknown
Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump · Scoped product and version facts
Version (version)Status (status)Version scheme (versionType)Exclusive end (lessThan)Inclusive end (lessThanOrEqual)Status changes (at → status)
Smiths Medical Medfusion 4000 Wireless Syringe Infusion PumpaffectedUnavailableUnavailableUnavailableNone listed