Texto simples

Texto da CNA em inglês

A Use of Hard-coded Credentials issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. The pump with default network configuration uses hard-coded credentials to automatically establish a wireless network connection. The pump will establish a wireless network connection even if the pump is Ethernet connected and active; however, if the wireless association is established and the Ethernet cable is attached, the pump does not attach the network stack to the wireless network. In this scenario, all network traffic is instead directed over the wired Ethernet connection.

Fatos de produtos e versões por escopo

Dados de correção desconhecidos; nenhuma versão corrigida é deduzida aqui. Dados ausentes não comprovam ausência de correção.

Exploração desconhecida; este catálogo não avalia exploração.

Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump

Fornecedor
n/a
Produto
Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump
Plataformas
Desconhecido
Status padrão
Desconhecido
Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump · Fatos de produtos e versões por escopo
Versão (version)Status (status)Esquema de versões (versionType)Fim exclusivo (lessThan)Fim inclusivo (lessThanOrEqual)Mudanças de status (at → status)
Smiths Medical Medfusion 4000 Wireless Syringe Infusion PumpaffectedIndisponívelIndisponívelIndisponívelNenhuma listada