CVE-2018-10631
CVE.org · Published
- Record status
- Published
- Identifier year (not publication year)
- 2018
- Source publication
- Source updated
- Imported by cveboard
- Source snapshot
- cve_2026-10-24_midnight_baseline ·
CNA source text (English)
Medtronic N'Vision Clinician Programmer Protection Mechanism Failure
The 8840 Clinician Programmer executes the application program from the 8870 Application Card. An attacker with physical access to an 8870 Application Card and sufficient technical capability can modify the contents of this card, including the binary executables. If modified to bypass protection mechanisms, this malicious code will be run when the card is inserted into an 8840 Clinician Programmer.
Scoped product and version facts
Fix information unknown; no fixed version is derived here. Missing information does not prove that no fix exists.
Exploitation unknown; this catalogue does not assess exploitation.
8840 N’Vision Clinician Programmer
- Vendor
- Medtronic
- Product
- 8840 N’Vision Clinician Programmer
- Platforms
- Unknown
- Default status
unaffected
| Version (version) | Status (status) | Version scheme (versionType) | Exclusive end (lessThan) | Inclusive end (lessThanOrEqual) | Status changes (at → status) |
|---|---|---|---|---|---|
| All versions | affected | Unavailable | Unavailable | Unavailable | None listed |
8870 N’Vision removable Application Card
- Vendor
- Medtronic
- Product
- 8870 N’Vision removable Application Card
- Platforms
- Unknown
- Default status
unaffected
| Version (version) | Status (status) | Version scheme (versionType) | Exclusive end (lessThan) | Inclusive end (lessThanOrEqual) | Status changes (at → status) |
|---|---|---|---|---|---|