cveboard RSS feed

CVE-2018-8868

CVE.org · Published

Historical CVEsJSONAI instructions

Record status
Published
Identifier year (not publication year)
2018
Source publication
Source updated
Imported by cveboard
Source snapshot
cve_2026-10-24_midnight_baseline ·

Plain text

CNA source text (English)

Medtronic MyCareLink Patient Monitor Exposed Dangerous Method or Function

Medtronic 24950 MyCareLink Monitor and 24952 MyCareLink Monitor contains debug code meant to test the functionality of the monitor's communication interfaces, including the interface between the monitor and implantable cardiac device. An attacker with physical access to the device can exploit other vulnerabilities to access this debug functionality. This debug functionality provides the ability to read and write arbitrary memory values to implantable cardiac devices via inductive or short range wireless protocols. An attacker with close physical proximity to a target implantable cardiac device can use this debug functionality.

Scoped product and version facts

Fix information unknown; no fixed version is derived here. Missing information does not prove that no fix exists.

Exploitation unknown; this catalogue does not assess exploitation.

24950 MyCareLink Monitor

Vendor
Medtronic
Product
24950 MyCareLink Monitor
Platforms
Unknown
Default status
unaffected
24950 MyCareLink Monitor · Scoped product and version facts
Version (version)Status (status)Version scheme (versionType)Exclusive end (lessThan)Inclusive end (lessThanOrEqual)Status changes (at → status)
All versionsaffectedUnavailableUnavailableUnavailableNone listed

24952 MyCareLink Monitor

Vendor
Medtronic
Product
24952 MyCareLink Monitor
Platforms
Unknown
Default status
unaffected
24952 MyCareLink Monitor · Scoped product and version facts
Version (version)Status (status)Version scheme (versionType)Exclusive end (lessThan)Inclusive end (lessThanOrEqual)Status changes (at → status)
All versionsaffectedUnavailableUnavailableUnavailableNone listed