Medtronic MiniMed 508 and Paradigm Series Insulin Pumps Improper Access Control

Medtronic MiniMed Insulin Pumps are designed to communicate using a wireless RF with other devices, such as blood glucose meters, glucose sensor transmitters, and CareLink USB devices. This wireless RF communication protocol does not properly implement authentication or authorization. An attacker with adjacent access to one of the affected insulin pump models can inject, replay, modify, and/or intercept data. This vulnerability could also allow attackers to change pump settings and control insulin delivery.

Scoped product and version facts

Fix information unknown; no fixed version is derived here. Missing information does not prove that no fix exists.

Exploitation unknown; this catalogue does not assess exploitation.

MiniMed 508 pump

Vendor
Medtronic
Product
MiniMed 508 pump
Platforms
Unknown
Default status
unaffected
MiniMed 508 pump · Scoped product and version facts
Version (version)Status (status)Version scheme (versionType)Exclusive end (lessThan)Inclusive end (lessThanOrEqual)Status changes (at → status)
All versionsaffectedUnavailableUnavailableUnavailableNone listed

MiniMed Paradigm 511 pump

Vendor
Medtronic
Product
MiniMed Paradigm 511 pump
Platforms
Unknown
Default status
unaffected
MiniMed Paradigm 511 pump · Scoped product and version facts
Version (version)Status (status)Version scheme (versionType)Exclusive end (lessThan)Inclusive end (lessThanOrEqual)Status changes (at → status)
All versionsaffectedUnavailableUnavailableUnavailableNone listed

MiniMed Paradigm 512/712 pumps

Vendor
Medtronic
Product
MiniMed Paradigm 512/712 pumps
Platforms
Unknown
Default status
unaffected
MiniMed Paradigm 512/712 pumps · Scoped product and version facts
Version (version)Status (status)Version scheme (versionType)Exclusive end (lessThan)Inclusive end (lessThanOrEqual)Status changes (at → status)
All versionsaffectedUnavailableUnavailableUnavailableNone listed

MiniMed Paradigm 712E pump

Vendor
Medtronic
Product
MiniMed Paradigm 712E pump
Platforms
Unknown
Default status
unaffected
MiniMed Paradigm 712E pump · Scoped product and version facts
Version (version)Status (status)Version scheme (versionType)Exclusive end (lessThan)Inclusive end (lessThanOrEqual)Status changes (at → status)
All versionsaffectedUnavailableUnavailableUnavailableNone listed

MiniMed Paradigm 515/715 pumps

Vendor
Medtronic
Product
MiniMed Paradigm 515/715 pumps
Platforms
Unknown
Default status
unaffected
MiniMed Paradigm 515/715 pumps · Scoped product and version facts
Version (version)Status (status)Version scheme (versionType)Exclusive end (lessThan)Inclusive end (lessThanOrEqual)Status changes (at → status)
All versionsaffectedUnavailableUnavailableUnavailableNone listed

MiniMed Paradigm 522/722 pumps

Vendor
Medtronic
Product
MiniMed Paradigm 522/722 pumps
Platforms
Unknown
Default status
unaffected
MiniMed Paradigm 522/722 pumps · Scoped product and version facts
Version (version)Status (status)Version scheme (versionType)Exclusive end (lessThan)Inclusive end (lessThanOrEqual)Status changes (at → status)
All versionsaffectedUnavailableUnavailableUnavailableNone listed

MiniMed Paradigm 522K/722K pumps

Vendor
Medtronic
Product
MiniMed Paradigm 522K/722K pumps
Platforms
Unknown
Default status
unaffected
MiniMed Paradigm 522K/722K pumps · Scoped product and version facts
Version (version)Status (status)Version scheme (versionType)Exclusive end (lessThan)Inclusive end (lessThanOrEqual)Status changes (at → status)
All versionsaffectedUnavailableUnavailableUnavailableNone listed

MiniMed Paradigm 523/723 pumps

Vendor
Medtronic
Product
MiniMed Paradigm 523/723 pumps
Platforms
Unknown
Default status
unaffected
MiniMed Paradigm 523/723 pumps · Scoped product and version facts
Version (version)Status (status)Version scheme (versionType)Exclusive end (lessThan)Inclusive end (lessThanOrEqual)Status changes (at → status)
0affectedcustomUnavailableSoftware Versions 2.4ANone listed

MiniMed Paradigm 523K/723K pumps

Vendor
Medtronic
Product
MiniMed Paradigm 523K/723K pumps
Platforms
Unknown
Default status
unaffected
MiniMed Paradigm 523K/723K pumps · Scoped product and version facts
Version (version)Status (status)Version scheme (versionType)Exclusive end (lessThan)Inclusive end (lessThanOrEqual)Status changes (at → status)
0affectedcustomUnavailableSoftware Versions 2.4ANone listed

MiniMed Paradigm Veo 554/754 pumps

Vendor
Medtronic
Product
MiniMed Paradigm Veo 554/754 pumps
Platforms
Unknown
Default status
unaffected
MiniMed Paradigm Veo 554/754 pumps · Scoped product and version facts
Version (version)Status (status)Version scheme (versionType)Exclusive end (lessThan)Inclusive end (lessThanOrEqual)Status changes (at → status)
0affectedcustomUnavailableSoftware Versions 2.6ANone listed

MiniMed Paradigm Veo 554CM/754CM pumps

Vendor
Medtronic
Product
MiniMed Paradigm Veo 554CM/754CM pumps
Platforms
Unknown
Default status
unaffected
MiniMed Paradigm Veo 554CM/754CM pumps · Scoped product and version facts
Version (version)Status (status)Version scheme (versionType)Exclusive end (lessThan)Inclusive end (lessThanOrEqual)Status changes (at → status)
0affectedcustomUnavailableSoftware Versions 2.7ANone listed