cveboard RSS feed

CVE-2024-48973

CVE.org · Published

Historical CVEsJSONAI instructions

Record status
Published
Identifier year (not publication year)
2024
Source publication
Source updated
Imported by cveboard
Source snapshot
cve_2026-10-24_midnight_baseline ·

Plain text

CNA source text (English)

Debug port on Life2000 Ventilator serial interface is enabled by default

The debug port on the ventilator's serial interface is enabled by default. This could allow an attacker to send and receive messages over the debug port (which are unencrypted; see 3.2.1) that result in unauthorized disclosure of information and/or have unintended impacts on device settings and performance.

Scoped product and version facts

Fix information unknown; no fixed version is derived here. Missing information does not prove that no fix exists.

Exploitation unknown; this catalogue does not assess exploitation.

Life2000 Ventilation System

Vendor
Baxter
Product
Life2000 Ventilation System
Platforms
Unknown
Default status
unaffected
Life2000 Ventilation System · Scoped product and version facts
Version (version)Status (status)Version scheme (versionType)Exclusive end (lessThan)Inclusive end (lessThanOrEqual)Status changes (at → status)
06.08.00.00 and prioraffectedUnavailableUnavailableUnavailableNone listed