cveboard RSS feed

CVE-2025-14096

CVE.org · Published

Historical CVEsJSONAI instructions

Record status
Published
Identifier year (not publication year)
2025
Source publication
Source updated
Imported by cveboard
Source snapshot
cve_2026-10-24_midnight_baseline ·

Plain text

CNA source text (English)

Credential Disclosure vulnerability in Radiometer Products

A vulnerability exists in multiple Radiometer products that allow an attacker with physical access to the analyzer possibility to extract credential information. The vulnerability is due to a weakness in the design and insufficient credential protection in operating system. Other related CVE's are CVE-2025-14095 & CVE-2025-14097. Affected customers have been informed about this vulnerability. This CVE is being published to provide transparency. Required Configuration for Exposure: Attacker requires physical access to the analyzer. Temporary work Around: Only authorized people can physically access the analyzer. Permanent solution: Local Radiometer representatives will contact all affected customers to discuss a permanent solution. Exploit Status: Researchers have provided a working proof-of-concept (PoC). Radiometer is not aware of any public exploit code at the time of this publication.

Scoped product and version facts

Fix information unknown; no fixed version is derived here. Missing information does not prove that no fix exists.

Exploitation unknown; this catalogue does not assess exploitation.

ABL90 FLEX and ABL90 FLEX PLUS Analyzers

Vendor
Radiometer Medical Aps
Product
ABL90 FLEX and ABL90 FLEX PLUS Analyzers
Platforms
  • Windows XP
  • Windows 7
Default status
unaffected
ABL90 FLEX and ABL90 FLEX PLUS Analyzers · Scoped product and version facts
Version (version)Status (status)Version scheme (versionType)Exclusive end (lessThan)Inclusive end (lessThanOrEqual)Status changes (at → status)
Windows 7 Operating systemaffectedRadiometer Customized Windows Operating SystemUnavailableUnavailableNone listed
Windows XP Operating systemaffectedRadiometer Customized Windows Operating SystemUnavailableUnavailableNone listed

AQT90 FLEX Analyzers

Vendor
Radiometer Medical Aps
Product
AQT90 FLEX Analyzers
Platforms
  • Windows 7
  • Windows XP
Default status
unaffected
AQT90 FLEX Analyzers · Scoped product and version facts
Version (version)Status (status)Version scheme (versionType)Exclusive end (lessThan)Inclusive end (lessThanOrEqual)Status changes (at → status)
Windows 7 Operating systemaffectedRadiometer Customized Windows Operating SystemUnavailableUnavailableNone listed
Windows XP Operating systemaffectedRadiometer Customized Windows Operating SystemUnavailableUnavailableNone listed

ABL800 BASIC and ABL800 FLEX analyzers

Vendor
Radiometer Medical Aps
Product
ABL800 BASIC and ABL800 FLEX analyzers
Platforms
  • Windows 7
  • Windows XP
Default status
unaffected
ABL800 BASIC and ABL800 FLEX analyzers · Scoped product and version facts
Version (version)Status (status)Version scheme (versionType)Exclusive end (lessThan)Inclusive end (lessThanOrEqual)Status changes (at → status)
Windows 7 Operating systemaffectedRadiometer Customized Windows Operating SystemUnavailableUnavailableNone listed
Windows XP Operating systemaffectedRadiometer Customized Windows Operating SystemUnavailableUnavailableNone listed