CVE-2025-14096
CVE.org · Publicado
- Status do registro
- Publicado
- Ano do identificador (não é o ano de publicação)
- 2025
- Publicação na fonte
- Atualização na fonte
- Importado pelo cveboard
- Snapshot da fonte
- cve_2026-10-24_midnight_baseline ·
Texto da CNA em inglês
Credential Disclosure vulnerability in Radiometer Products
A vulnerability exists in multiple Radiometer products that allow an attacker with physical access to the analyzer possibility to extract credential information. The vulnerability is due to a weakness in the design and insufficient credential protection in operating system. Other related CVE's are CVE-2025-14095 & CVE-2025-14097. Affected customers have been informed about this vulnerability. This CVE is being published to provide transparency. Required Configuration for Exposure: Attacker requires physical access to the analyzer. Temporary work Around: Only authorized people can physically access the analyzer. Permanent solution: Local Radiometer representatives will contact all affected customers to discuss a permanent solution. Exploit Status: Researchers have provided a working proof-of-concept (PoC). Radiometer is not aware of any public exploit code at the time of this publication.
Fatos de produtos e versões por escopo
Dados de correção desconhecidos; nenhuma versão corrigida é deduzida aqui. Dados ausentes não comprovam ausência de correção.
Exploração desconhecida; este catálogo não avalia exploração.
ABL90 FLEX and ABL90 FLEX PLUS Analyzers
- Fornecedor
- Radiometer Medical Aps
- Produto
- ABL90 FLEX and ABL90 FLEX PLUS Analyzers
- Plataformas
- Windows XP
- Windows 7
- Status padrão
unaffected
| Versão (version) | Status (status) | Esquema de versões (versionType) | Fim exclusivo (lessThan) | Fim inclusivo (lessThanOrEqual) | Mudanças de status (at → status) |
|---|---|---|---|---|---|
| Windows 7 Operating system | affected | Radiometer Customized Windows Operating System | Indisponível | Indisponível | Nenhuma listada |
| Windows XP Operating system | affected | Radiometer Customized Windows Operating System | Indisponível | Indisponível | Nenhuma listada |
AQT90 FLEX Analyzers
- Fornecedor
- Radiometer Medical Aps
- Produto
- AQT90 FLEX Analyzers
- Plataformas
- Windows 7
- Windows XP
- Status padrão
unaffected
| Versão (version) | Status (status) | Esquema de versões (versionType) | Fim exclusivo (lessThan) | Fim inclusivo (lessThanOrEqual) | Mudanças de status (at → status) |
|---|---|---|---|---|---|
| Windows 7 Operating system | affected | Radiometer Customized Windows Operating System | Indisponível | Indisponível | Nenhuma listada |
| Windows XP Operating system | affected | Radiometer Customized Windows Operating System | Indisponível | Indisponível | Nenhuma listada |
ABL800 BASIC and ABL800 FLEX analyzers
- Fornecedor
- Radiometer Medical Aps
- Produto
- ABL800 BASIC and ABL800 FLEX analyzers
- Plataformas
- Windows 7
- Windows XP
- Status padrão
unaffected
| Versão (version) | Status (status) | Esquema de versões (versionType) | Fim exclusivo (lessThan) | Fim inclusivo (lessThanOrEqual) | Mudanças de status (at → status) |
|---|---|---|---|---|---|
| Windows 7 Operating system | affected | Radiometer Customized Windows Operating System | Indisponível | Indisponível | Nenhuma listada |
| Windows XP Operating system | affected | Radiometer Customized Windows Operating System | Indisponível | Indisponível | Nenhuma listada |