Remote Code Execution Vulnerability in Radiometer Products
A vulnerability in the application software of multiple Radiometer products may allow remote code execution and unauthorized device management when specific internal conditions are met. Exploitation requires that a remote connection is established with additional information obtained through other means. The issue is caused by a weakness in the analyzer’s application software. Other related CVE's are CVE-2025-14095 & CVE-2025-14096. Affected customers have been informed about this vulnerability. This CVE is being published to provide transparency. Required Configuration for Exposure: Affected application software version is in use and remote support feature is enabled in the analyzer. Temporary work Around: If the network is not considered secure, please remove the analyzer from the network. Permanent solution: Customers should ensure the following: • The network is secure, and access follows best practices. Local Radiometer representatives will contact all affected customers to discuss a permanent solution. Exploit Status: Researchers have provided working proof-of-concept (PoC). Radiometer is not aware of any publicly available exploits at the time of this publication.
Scoped product and version facts
Fix information unknown; no fixed version is derived here. Missing information does not prove that no fix exists.
Exploitation unknown; this catalogue does not assess exploitation.
ABL90 FLEX and ABL90 FLEX PLUS Analyzers
- Vendor
- Radiometer Medical Aps
- Product
- ABL90 FLEX and ABL90 FLEX PLUS Analyzers
- Platforms
- Unknown
- Default status
unaffected
| Version (version) | Status (status) | Version scheme (versionType) | Exclusive end (lessThan) | Inclusive end (lessThanOrEqual) | Status changes (at → status) |
|---|---|---|---|---|---|
| Application software versions < 3.5MR11 with Windows 7, Windows XP operating systems | affected | ABL90 DMS (Data Management system) | Unavailable | Unavailable | None listed |
| Application software versions >= 3.5MR11 with Windows 10 operating system | unaffected | ABL90 DMS (Data Management system) | Unavailable | Unavailable | None listed |
ABL800 BASIC and ABL800 FLEX Analyzers
- Vendor
- Radiometer Medical Aps
- Product
- ABL800 BASIC and ABL800 FLEX Analyzers
- Platforms
- Unknown
- Default status
unaffected
| Version (version) | Status (status) | Version scheme (versionType) | Exclusive end (lessThan) | Inclusive end (lessThanOrEqual) | Status changes (at → status) |
|---|---|---|---|---|---|
| Application software versions < 6.20MR2 with Windows 7, Windows XP operating systems | affected | ABL800 DMS(Data Management System) | Unavailable | Unavailable | None listed |
| Application software versions >= 6.20MR2 with Windows 10 operating system | unaffected | ABL800 DMS(Data Management System) | Unavailable | Unavailable | None listed |
AQT90 FLEX Analyzers
- Vendor
- Radiometer Medical Aps
- Product
- AQT90 FLEX Analyzers
- Platforms
- Unknown
- Default status
unaffected
| Version (version) | Status (status) | Version scheme (versionType) | Exclusive end (lessThan) | Inclusive end (lessThanOrEqual) | Status changes (at → status) |
|---|---|---|---|---|---|
| Application software versions <= 8.13 MR2 with Windows 7, Windows XP operating systems | affected | AQT90 DMS (Data Management System) | Unavailable | Unavailable | None listed |
| Application software versions >= 8.13 MR2 with Windows 10 operating system | unaffected | AQT90 DMS (Data Management System) | Unavailable | Unavailable | None listed |