cveboard RSS feed

CVE-2025-14097

CVE.org · Published

Historical CVEsJSONAI instructions

Record status
Published
Identifier year (not publication year)
2025
Source publication
Source updated
Imported by cveboard
Source snapshot
cve_2026-10-24_midnight_baseline ·

Plain text

CNA source text (English)

Remote Code Execution Vulnerability in Radiometer Products

A vulnerability in the application software of multiple Radiometer products may allow remote code execution and unauthorized device management when specific internal conditions are met. Exploitation requires that a remote connection is established with additional information obtained through other means. The issue is caused by a weakness in the analyzer’s application software.                                                                                                                                                                                                Other related CVE's are CVE-2025-14095 & CVE-2025-14096.                                                                                                      Affected customers have been informed about this vulnerability. This CVE is being published to provide transparency. Required Configuration for Exposure: Affected application software version is in use and remote support feature is enabled in the analyzer.                                                                                                                                                                        Temporary work Around: If the network is not considered secure, please remove the analyzer from the network.                         Permanent solution: Customers should ensure the following: • The network is secure, and access follows best practices. Local Radiometer representatives will contact all affected customers to discuss a permanent solution.                                                      Exploit Status: Researchers have provided working proof-of-concept (PoC). Radiometer is not aware of any publicly available exploits at the time of this publication.

Scoped product and version facts

Fix information unknown; no fixed version is derived here. Missing information does not prove that no fix exists.

Exploitation unknown; this catalogue does not assess exploitation.

ABL90 FLEX and ABL90 FLEX PLUS Analyzers

Vendor
Radiometer Medical Aps
Product
ABL90 FLEX and ABL90 FLEX PLUS Analyzers
Platforms
Unknown
Default status
unaffected
ABL90 FLEX and ABL90 FLEX PLUS Analyzers · Scoped product and version facts
Version (version)Status (status)Version scheme (versionType)Exclusive end (lessThan)Inclusive end (lessThanOrEqual)Status changes (at → status)
Application software versions < 3.5MR11 with Windows 7, Windows XP operating systemsaffectedABL90 DMS (Data Management system)UnavailableUnavailableNone listed
Application software versions >= 3.5MR11 with Windows 10 operating systemunaffectedABL90 DMS (Data Management system)UnavailableUnavailableNone listed

ABL800 BASIC and ABL800 FLEX Analyzers

Vendor
Radiometer Medical Aps
Product
ABL800 BASIC and ABL800 FLEX Analyzers
Platforms
Unknown
Default status
unaffected
ABL800 BASIC and ABL800 FLEX Analyzers · Scoped product and version facts
Version (version)Status (status)Version scheme (versionType)Exclusive end (lessThan)Inclusive end (lessThanOrEqual)Status changes (at → status)
Application software versions < 6.20MR2 with Windows 7, Windows XP operating systemsaffectedABL800 DMS(Data Management System)UnavailableUnavailableNone listed
Application software versions >= 6.20MR2 with Windows 10 operating systemunaffectedABL800 DMS(Data Management System)UnavailableUnavailableNone listed

AQT90 FLEX Analyzers

Vendor
Radiometer Medical Aps
Product
AQT90 FLEX Analyzers
Platforms
Unknown
Default status
unaffected
AQT90 FLEX Analyzers · Scoped product and version facts
Version (version)Status (status)Version scheme (versionType)Exclusive end (lessThan)Inclusive end (lessThanOrEqual)Status changes (at → status)
Application software versions <= 8.13 MR2 with Windows 7, Windows XP operating systemsaffectedAQT90 DMS (Data Management System)UnavailableUnavailableNone listed
Application software versions >= 8.13 MR2 with Windows 10 operating systemunaffectedAQT90 DMS (Data Management System)UnavailableUnavailableNone listed